Good technique when working within a Class II Biological Safety Cabinet (BSC) will minimise air turbulence and prevent splatter or unwanted spread of aerosols. Here are some tips for good technique that will maximise potential protection of personnel, product and environment.

Tip 1: Know your Airflow Biological Safety Cabinets provide personnel, product, and environmental protection through the use of HEPA Filtered air. Knowing how the air is passed through the cabinet an essential piece of the puzzle to know.

Tip 2: Work at the Proper Sash Level Proper airflows are reached by a certifier measuring air speed on an annual basis (or more) to put a stamp of approval that the cabinet is fully operation to provide personnel, product, and environmental protection.

Tip 3: Never Cover the Air Grill Covering the air grill at the front of the cabinet compromises airflow integrity; blocking the grill can cause laboratory air to enter the work zone or even work zone air to enter the laboratory environment.

Tip 4: Minimise Movement The rapid movement of a worker’s arms in a sweeping motion into and out of the cabinet will disrupt the air curtain and may compromise the partial barrier containment provided by the BSC. Moving arms in and out slowly, perpendicular to the face opening of the cabinet will reduce risk. Other personnel activities in the room (e.g. rapid movement, open/closing room doors etc.) may also disrupt the cabinet air barrier.

Tip 5: Reduce Splatter Many common procedures conducted in Biological Safety Cabinets may create splatter or aerosols. Good microbiological techniques should always be used when working in a Biological Safety Cabinet to minimis e this splatter, aerosols, and the-potential for personnel exposure to infectious materials manipulated within the cabinet.

Tip 6: Know Your Work Area The middle third of the work surface is the ideal area to be used. All operations should be performed at least 4 inches from the front grille on the work surface.  Materials or equipment placed inside the cabinet may cause disruption to the airflow, resulting in turbulence, possible cross-contamination, and/or breach of containment.

Tip 7: Work form Clean to Contaminated Active work should flow from the clean to contaminated area across the work surface. Materials and supplies should be placed in such a way as to limit the movement of 'dirty' items over 'clean' ones. All materials should be placed toward the rear edge of the work surface and away from the front grille of the cabinet to take advantage of the air split.

Tip 8: Working with Tubes Open tubes or bottles should NOT be held in a vertical position. Bottle or tube caps should not be placed on the toweling. Items should be recapped or covered as soon as possible.

Tip 9: Working with Petri Dishes Investigators working with Petri dishes and tissue culture plates should hold the lid above the open sterile surface to minimise direct impaction of downward air.

Tip 10: Working with Aspirator Bottles or Suction Flasks Aspirator bottles or suction flasks should be connected to an overflow collection flask containing appropriate disinfectant, and to an in-line HEPA or equivalent filter.

Visit the website to learn more.

For laboratories looking for a relatively small autoclave but cannot accommodate a benchtop model and there is limited floor space, a vertical chamber steril...

Microtrac MRB’s BELSORP MINI X allows measuring specific surface area, pore size distribution and total pore volume. It shows outstanding features resultin...

PCR applications have been in the spotlight for over two years with the arrival of Covid-19 and PCR being an essential part of the testing system around the...

The importance of measurement cell cleanliness Proper cleaning of a density meter's measurement cell is vital in achieving the most precise results possib...

In This Edition Chromatography - Automated Sample Preparation:The Missing Hyphen to Hypernation - New Low Volume Air Sampler for PFAS Analysis - Analytical Intelligence Starts with the Samp...

ACS National Meeting & Expo, Fall 2022

Aug 21 2022 Chicago, IL, USA & Online

International Labmate Limited Oak Court Business Centre Sandridge Park, Porters Wood St Albans Hertfordshire AL3 6PH United Kingdom

T +44 (0)1727 858 840 F +44 (0)1727 840 310 E info@labmate-online.com

Copyright © 2022 Labmate Online. All rights reserved.

200)" class="scrollToTop">Top